Features:

  • Two Cut Modes, Cut I & Cut II
  • Blend with 10 settings
  • Presets
  • Two levels of coagulation: Pinpoint and Spray
  • Digitally controlled cutting and coagulation delivers consistent, repeatable power into varying load impedances, with FDFS™ (fast digital feedback system) -or- FCFS™ (first come, first served)
  • Return electrode sensing and contact quality monitoring
  • Split or soid return electrodes
  • NEM™ (neutral electrode monitoring)
  • Defibrillator-proof type CF equipment
  • Color coordinated controls
  • Pre-Set buttons to save 10 physician settings
  • Mono & Bipolar with Footswitch
  • Footswitch indicator on face of unit
  • Dual voltage 50/60 Hz
  • Does not require calibration
  • Remembers last used settings at power on

Optional Accessories:

  • BiPolar Footswitch required for BiPolar operation.
  • Monopolar Footswitches
  • Pens / Pencils (Mono and Bipolar)
  • Foreceps (Mono and Bipolar)
  • Grounding Cord/Plate/Pads
Warranty: 2 Years

Specifications

  • Pre-Set buttons to save 10 physician settings.
  • Duty Cycle: 300 watt @ 300 ohms
  • Two year warranty

Dimensions and Weight:

  • Width: 31.1 cm (12.25 in.)
  • Height: 15.3 cm (6.00 in.)
  • Depth: 41.3 cm (16.25 in.)
  • Weight: <8.75 kg (<19 lbs)

Input Power:

  • Input Voltage: 100-240 VAC ± 10%
  • Mains line frequency range (nominal): 50/60 HZ
  • Power Consumption: 500 VA
  • Fuses (two): 6.3 A (slow blow)
  • Max Wattage: 300 watts

Monopolar:

  • Cut I - Max W:300 @ 300 ohms, kHz:490 ± 5
  • Cut II - Max W:300 @300 ohms, kHz:490 ± 5
  • Blend (Max) - Max W:200 @ 300 ohms, kHz:490 ± 5
  • Pinpoint - Max W:120 @ 500 ohms, kHz: 490 ± 5
  • Spray - Max W: 80 @ 500 ohms, kHz: 490 ± 5

BiPolar:

  • Max W: 80 @ 1500 ohms
  • kHz:490 ± 5