DRE NeoGlider Transport Incubator

Ask About Financing

Available in US only.

Need Equipment Advice?

Ask Our Experts! (800) 462-8195

Details

Features

  • Use AC power and DC power alternatively, and DC 12V or DC 24V power in an ambulance
  • Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately
  • Can be equipped with >37 C temperature set function
  • Failure alarm indication
  • Double wall hood with side door for pulling out the bassinet
  • Natural air flow humidity
  • Adjustable height

Configurations available with

  • Main body, including the hood, controller, bassinet, internal batter and observation lamp
  • Oxygen cylinders
  • Oxygen supply system
  • Skin temperature sensor
  • IV pole
  • Mattress
  • Adjustable stand
  • >37 C override temperature mode (optional)

Specifications

General

  • AC power supply: AC110V - 240V/50-60Hz
  • DC power supply: DC12V/10A or DC24V/6A
  • Input power: 400VA
  • Air-temperature control range: 25 C - 37 C
  • Override mode: 37 C - 38 C
  • Baby temperature control range: 34 C - 37 C
  • Override mode: 37 C - 37.5 C
  • Skin temperature sensor accuracy: ≤0.3 C
  • Temperature variability: ≤1.0 C
  • Temperature uniformity: ≤1.5 C (mattress is on the horizontal position)
  • Noise inside hood: ≤55dB(A)
  • Environment noise ≤45dB(A)
  • Failure alarm: Over-temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure alarm
  • Operating at temperature equilibrium: 90 min (Set temp: 36 C | Ambient temperature: 15 C)

Environmental Temperature

  • Operating range: 10 C ~ 30 C
  • Set temperature should be 3 C higher than the environment temperature
  • Ambient air movement rate: ≤1.0 m/s

Package

  • Gross Weight:
    • Main unit: 80 kg
    • Cabinet: 53 kg

Size

  • Main body: 109 cm x 65.5 cm x 76 cm
  • Cabinet: 138 cm x 61 cm x 40 cm