Refurbished Siemens 300 & 300A Servo Ventilators offer a wide range of ventilation modes together with an Automode function for greater treatment possibilities, shorter weaning times, and increased patient comfort.

Features:

  • Neonatal to adult patient range
  • Wide range of ventilation modes
  • Controlled, Support, or Combined ventilation modes
  • Easy to use
  • Optional graphical display, and compressor
  • Refurbished

Configurable Model:

  • Siemens 300A Servo Ventilator - #40300A
    • Automode for fully adaptive patient interaction
    • Neonate to adult patient range
    • Wide range of ventilation modes: Controlled, Supported, Combined
    • Easy to use
    • Refurbished

Available Accessories For Product:

  • Carts
  • Breathing Systems
  • Compressors
  • External Power Supply
  • Humidifiers

Product Options

Product Name ID Price
Siemens 300A Servo Ventilator Siemens 300A Servo Ventilator 380 Get Quote
Siemens 300 Servo Ventilator Siemens 300 Servo Ventilator 1430 Get Quote

Specifications

Dimensions:

  • Patient unit: 9.5" (W) x 14.6" (D) x 9.4" (H)
    User interface: 17.0" (W) x 15.9" (D) x 12.8" (H)

Weight: Approx. 52.9 lbs

Ventilation Modes:

  • Controlled
  • Supported
  • Combined

Patient Range:

  • Large/Medium/Small Animal

Classification:

  • IEC60601-1/EN60 601-1 Type B

Method of Triggering:

  • Flow and Pressure

Flow Range: 0.1 ml/s - 3 l/s

Power Supply: 100, 120, 220, and 240 VAC 10%, 50-60Hz

Alarms:

  • Airway Pressure (Upper): 15-120 cm H2O
  • High Continuous Pressure: Set PEEP-level +15 cm H2O for more than 15 seconds.
  • O2 concentration: Set value 6%. Lower alarm limit cannot go below 18% or exceed 90%.
  • Expired Minute Volume (Upper Alarm Limit): Adult/Pediatric Range: 0-60 l/min, Neonate Range: 0-6 l/min.
  • Expired Minute Volume (Lower Alarm Limit): Adult/Pediatric Range: 0.3-40 l/min, Neonate Range: 0.06-4 l/min.
  • Apnea: Adult: 20 s. Pediatric: 15 s. Neonates: 10 s.

Monitoring:

  • Frequency: CMV or SIMV frequency 1% or 0.5 b/min.
  • Pressures: Peak, Pause, Mean and End exp. 5% b/min or 2 cm H2O
  • Insp. Tidal Volume: Adult Range:50-3999 ml 6% or 3 ml. Pediatric Range: 10-399 ml 6% or 0.5 ml. Neonate Range: 2.0-39 ml 6% or 0.5 ml.
  • Exp. Tidal Volume: Adult Range: 50-3999 ml 6% or 3 ml. Pediatric Range: 10-399 ml 6% or 0.5 ml. Neonate Range: 2.0-39 ml 6% or 0.5 ml.
  • Exp. Minute Volume: Adult Range: 4.0-60 l/min 6% or 0.1 l/min. Pediatric Range: 1.0-5.0 l/min 6% or 0.05 l/min. Neonate Range: 0.20-1.50 l/min 6% or +/-0.02 l/min.
  • Flow Rate: 0.00-3.00 l/s or 0.00-180 l/min 1% or 1% or 1 digit.
  • O2 Concentration: 5% of read value