Baxter Infus O.R. Syringe Pump - Refurbished

Ask About Financing

Available in US only.

Need Equipment Advice?

Ask Our Experts! (800) 462-8195

Details

FEATURES:

  • Modify dose without multiple reprogramming steps.
  • Adjust settings easily at any time during infusion.
  • Change dial settings with a single hand for an immediate response.
  • Save calculation time.
  • Minimize risk of calculation error.
  • Change Infusion Parameters/Rates with SMART LABELS
  • SMART LABELS are magnetically coded/drug specific.
  • Select from the available SMART Label library.
  • Ideal for administering drugs requiring multiple dosage adjustments.

Specifications

  • Size - 9.2 x 4.5 x 2.0 in
  • Weight – 2 lbs (with batteries)
  • Syringes - Monoject or B-D; Monoject or B-D 20 cc
  • Accuracy – Infusion: linear rate 3%. Bolus: linear displacement 3%
  • Batteries – Four “C” Alkaline Cells
  • Battery Life – Typical Use: 150 hours; at low battery and typical use: 4 hours
  • Operating Voltage – 5-7 volts
  • Low Battery Voltage – 5.0 Volts (approx.)
  • Flow Rates – 0-600 mL/hr
  • Flow Profile – Bolus: Continuous; Infusion: Pulsed Intermittently
  • Maximum Occlusion Pressure – 11 psi with 60 cc syringe; 20 psi with 20 cc syringe
  • Occlusion Detection Time – 120 seconds at 36 mL/hr (inverse related to flow rate)
  • Volume Stored on Occlusion – 1.1 mL (approx.)
  • Back Pressure Effect on Accuracy – None to Occlusion Pressure