Measurement Range
- SpO2 - 0 – 100%
- SpMet - 0 – 99.9%
- SpCO - 0 – 99%
- SpHb - 0 – 25 g/dL
- SpOC - 0 – 35 ml of O2/dl of blood
- Pulse Rate - 25 – 240 bpm
- Perfusion Index - 0.02 – 20%
- PVI - 0 – 100%
Oxygen Saturation Accuracy: SpO2
- Saturation - 60 – 80%
- No Motion
- Adults/Infants/Pediatrics - + 2%
- Neonates - + 3%
- Motion
- Adults/Infants/Pediatrics/Neonates - + 3%
- Low Perfusion
- Adults/Infants/Pediatrics/Neonates - + 2%
Pulse Rate Accuracy
- Pulse Rate - 25 – 240 bpm
- No Motion
- Adults/Infants/Pediatrics/Neonates - + 3 bpm
- Motion
- Adults/Infants/Pediatrics/Neonates - + 5 bpm
- Low Perfusion
- Adults/Infants/Pediatrics/Neonates v + 3 bpm
Carboxyhemoglobin Saturation Accuracy (%SpCO)*
- Adults/Infants/Pediatrics - 1 – 40% + 3%
Methemoglobin Saturation Accuracy (%SpMet)*
- Adults/Infants/Pediatrics/Neonates - 1 – 15% + 1%
Total Hemoglobin Accuracy (SpHb g/dL)
- Adults/Pediatrics - 8 – 17 g/dL + 1 g/dL
Resolution
- Oxyhemoglobin Saturation (%SpO2) - 1%
- Carboxyhemoglobin Saturation (%SpCO), digital display - 1%
- Methemoglobin Saturation (%SpMet), digital display - 0.1%
- Total Hemoglobin (SpHb g/dL) - 0.1 g/dL
- Pulse Rate (bpm): - 1 bpm
Modes
- Averaging mode - 2, 4, 8, 10, 12, 14, or 16 seconds
- Sensitivity - APOD, FastSat, Normal, and Max
Alarms
- High/low audible and visual alarms for parameters (SpO2 range
- 1 – 99% then “---”, SpHb range 0.1 – 24.5 g/dL then “---” SpCO, range
- 1 – 99% then “---”, SpMet range 1 – 99% then “---”, pulse rate range
- 25 – 240 bpm), sensor condition, system failure and low battery alarms
- Alarm volume range - 45 – 85 db
Display/Indicators
- Data display: %SpO2, %SpMet, %SpCO, SpHb g/dL, SpOC ml/dl, PVI,wireless, sensitivity, patient status light, device profi le light, pulse rate,alarm status, alarm silenced status, AC power, Signal IQ / pleth bar,perfusion index bar, battery status, no sensor, sensor off
- Display Language - English (default)
- APOD, Normal, and Max - LED